Functions
  
 Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
 1)ECG Heart rate(HR)
 ECG waveform
 Arrhythmia and ST-segment analysis
 2)RESP Respiration rate(RR)
 Respiration waveform
 3)SpO2 Oxygen saturation(SpO2)
 Plethysmogram(PLETH) waveform
 Pulse rate(PR)
 Bar graph
 4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
 5)TEMP T1, T2, TD
 It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
 
 Features
 1)8'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).
 2)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
 3)All-round monitor for adult, pediatric and neonate.
 4)With standard interface, OxyCRG, trend chard, big character interface and view bed, convenient to observe.
 5)Waveform, parameter color and location can be set optionally.
 6)Operation mode: keys and buttons.
 7)Display up to 8-channel waveform on one screen.
 8)Display 7-lead ECG waveform on one screen, and with ECG step function.
 9)Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
 10)HRV analysis function.
 11)NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
 12)Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
 13)Drug concentration calculation and titration table functions.
 14)Print trend table data with one-key.
 15)Connect to Central Monitoring System by 3G, WiFi or wired mode.
 16)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
 17)Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
 
 Performance
  
 ECG
 Lead mode: 3-lead or 5-lead
 Lead selection: I, II, III, aVR, aVL, aVF, V
 Waveform: 5-lead, 2-channel
 3-lead, 1-channel
 Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
 Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
 HR:
 Measurement and alarm range: 15~350bpm
 Accuracy: ±1% or ±1bpm, whichever is greater
 Alarm accuracy: ± 2 bpm
 Resolution: 1 bpm
 ST-segment monitoring:
 Measurement and alarm range: -2.0mV~+2.0mV
 Accuracy: -0.8mV~+0.8mV ±0.04mv or ±10%, whichever is greater
 Other range: unspecified
 Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
 Pacemaker: yes
 RESPIRATION
 Method: R-F(RA-LL) Impedance
 Respiration rate:
 Measurement and alarm range: 0~150rpm
 Resolution: 1 rpm
 Measurement accuracy: ±2 rpm
 Alarm accuracy: ±3 rpm
 Apnea alarm: 10~40s
 Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
 NIBP
 Method: Oscillometry
 Mode: Manual/AUTO/STAT
 Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
 Measurement period in STAT mode: 5 minutes
 Measurement and alarm range: 10 ~ 270 mmHg
 Resolution: 1 mmHg
 Cuff pressure accuracy: ±3 mmHg
 Measurement accuracy:
 Maximal mean deviation: ±5 mmHg
 Maximal standard deviation: 8mmHg
 Over-pressure protection:
 Adult mode: 297±3 mmHg
 Pediatric mode: 240±3 mmHg
 Neonatal mode: 147±3 mmHg
 SpO2
 Measurement and alarm range: 0 ~ 100%
 Resolution: 1%
 Measurement accuracy: 70%~100%: ±2%;
 0%~69%: unspecified
 PR
 Measurement and alarm range: 30 ~ 250 bpm
 Measurement accuracy: ±2 bpm or ±2%, whichever is greater
 TEMP
 Channel: dual-channel
 Measurement and alarm range: 0 ~ 50℃
 Resolution: 0.1 ℃
 Accuracy: ±0.1 ℃
 EtCO2
 Method: Sidestream or Mainstream
 Measurement range for CO2: 0 ~ 150mmHg
 Resolution for CO2:
 0.1 mm Hg: 0 ~ 69 mm Hg
 0.25 mm Hg: 70 ~ 150 mm Hg
 Accuracy for CO2:
 0 ~ 40 mm Hg: ±2 mm Hg
 41 ~ 70 mm Hg: ±5%
 71 ~ 100 mm Hg: ±8%
 101 ~ 150 mm Hg: ±10%
 AwRR range: 2~150 rpm
 AwRR accuracy: ±1 rpm
 Apnea alarm: yes
 IBP
 Channel: dual-channel
 Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
 Measurement and alarm range: -10~300 mm Hg
 Resolution: 1 mmHg
 Accuracy: ±2% or 1mm Hg, whichever is greater
 Power supply:100-240V~, 50/60Hz
 Safety classification: Class I, type CF defibrillation-proof applied part
 
 Accessories
  
 Adult fingertip SpO2 probe (5-pin)
 Adult NIBP cuff
 NIBP extension tube
 ECG lead cable
 ECG electrode
 Temperature probe
 Power adapter
 Power cord
 User Manual
  
 Physical characteristic
  
 Dimension: 235 mm(L) × 141 mm(W) × 222 mm(H)